ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2013-07802
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Date of Event
- May 1, 2013
- Report Date
- May 2, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE FIELD REPRESENTATIVE SPOKE WITH THE CLINIC. IT WAS DECIDED TO OBSERVE FOR THE MOMENT. NO IN-CLINIC FOLLOW-UP WAS PLANNED AND A REMOTE FOLLOW-UP WAS SCHEDULED IN ONE MONTH TO RE-EVALUATE. IT WAS REPORTED THAT SHOCK IMPEDANCE MEASUREMENTS HAD BEEN BETWEEN 95-117 OHMS. NO ADVERSE PATIENT EFFECTS WERE REPORTED. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND ASSOCIATED RIGHT VENTRICULAR (RV) LEAD EXHIBITED A HIGH, OUT-OF-RANGE SHOCK IMPEDANCE MEASUREMENT. AN ALERT WAS ISSUED IN THE PATIENT'S REMOTE MONITORING SYSTEM FOR THIS MEASUREMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310611 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0181 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR | 0181| E102 |