FDA Adverse Event Malfunction Summary report: N

EASYTRAK 3

MDR report key: 3211055 · Received July 8, 2013

Report

Report Number
2124215-2013-08322
Event Type
Malfunction
Date Received
July 8, 2013
Date of Event
May 1, 2013
Report Date
June 10, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD WAS ATTEMPTED AT IMPLANT. IT WAS REPORTED THAT PACING WAS NOT DELIVERED WHEN EXPECTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311178 EASYTRAK 3 IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4549

Patients

Seq Age Sex Outcome Treatment
1