EASYTRAK 3
Report
- Report Number
- 2124215-2013-08322
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Date of Event
- May 1, 2013
- Report Date
- June 10, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- OJX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- PHYSICIAN
Narratives
THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD WAS ATTEMPTED AT IMPLANT. IT WAS REPORTED THAT PACING WAS NOT DELIVERED WHEN EXPECTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.