FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 3211042 · Received July 8, 2013

Report

Report Number
2124215-2013-08369
Event Type
Injury
Date Received
July 8, 2013
Date of Event
May 2, 2013
Report Date
May 2, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING ROUTINE TESTING, THIS RIGHT VENTRICULAR (RV) LEAD MEASURED A SHOCK IMPEDANCE OF ABOUT 170 OHMS. ADDITIONALLY, THE PACE IMPEDANCE MEASUREMENT WAS GREATER THAN 2000 OHMS. ALL OTHER PARAMETERS ARE NORMAL AND THERE WAS NO NOISE ON THE SHOCK CHANNEL. A REVISION PROCEDURE WAS PERFORMED 11 DAYS LATER, AND A DISCONNECTED DISTAL COIL WAS REPORTED. THE SOURCE OF THE OUT OF RANGE IMPEDANCE MEASUREMENTS WAS DUE TO A CONNECTION ISSUE WITH THE PULSE GENERATOR. ALL PRODUCTS WERE SUCCESSFULLY RECONNECTED WITH NORMAL MEASUREMENTS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310018 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI - DEL CARIBE 0175

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R