ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2013-07836
- Event Type
- Injury
- Date Received
- July 8, 2013
- Date of Event
- April 30, 2013
- Report Date
- May 16, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY THE LEAD WAS RETURNED SEVERED 138 MM FROM THE TERMINAL PIN, TWO SEGMENTS WERE RETURNED, A TERMINAL END SEGMENT AND A TIP SEGMENT. THE TOTAL LENGTH WAS 640 MM. ALL FILARS APPEAR CUT. SET SCREW MARKS WERE NOTED, THREE ON THE IS-1 TERMINAL PIN, TWO ON THE DISTAL HIGH VOLTAGE TERMINAL END, AND TWO ON THE PROXIMAL HIGH VOLTAGE TERMINAL END. THE LEAD WAS RETURNED WITH THE EXTRACTING STYLET STUCK IN THE TIP SEGMENT OF THE LEAD. A SUTURE WAS TIED TIGHT AROUND THE LEAD BODY OF THE TIP SEGMENT FOR REMOVAL PURPOSES. SUTURE SLEEVE INDENTS WERE NOTED IN THE INSULATION 203 AND 210 MM FROM THE TERMINAL PIN. A CUT WAS NOTED IN THE INSULATION OF THE TIP SEGMENT OF THE LEAD. LASER DAMAGE WAS NOTED IN THE INSULATION OF THE TIP SEGMENT OF THE LEAD. THE TISSUE WAS NOTED ON THE LEAD BODY IN SEVERAL PLACES. SOME CALCIFICATION WAS NOTED AS WELL. BLOOD AND BODY FLUID WAS NOTED IN THE RATE SENSE LUMEN FROM THE TERMINAL PIN TO THE TIP AND IN THE HELIX HOUSING. THE HELIX WAS EXTENDED. THE LEAD PASSED CONTINUITY TESTING WITH MANIPULATION.
THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.
--
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD REVISION WAS SCHEDULED DUE TO GREATER THAN 200 OHM IMPEDANCE MEASUREMENTS. DURING THE REVISION PROCEDURE THE LEAD WAS SUCCESSFULLY REPLACED. DURING FINAL TESTING, THE RIGHT ATRIAL (RA) LEAD WAS RECONNECTED TO THE DEVICE HOWEVER, NOISE AND GREATER THAN 2000 OHM PACING IMPEDANCE MEASUREMENTS WERE OBSERVED (THAT WERE NOT PRESENT PRIOR TO THE PROCEDURE). TESTING THROUGH THE PACING SYSTEM ANALYZER (PSA) REVEALED NORMAL VALUES WITH NO NOISE. THE DEVICE WAS EXPLANTED AND A NEW DEVICE WAS SUCCESSFULLY IMPLANTED. NORMAL MEASUREMENTS WERE NOTED WITH THE NEW DEVICE. THE PREVIOUS DEVICE AND RV LEAD WERE RETURNED FOR ANALYSIS TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310887 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0185 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Hospitalization| L| R | (B)(4)| 4136| N118| 0185| 0181| N164| 4518 |