FDA Adverse Event Injury Summary report: N

FINELINE II

MDR report key: 3210998 · Received July 8, 2013

Report

Report Number
2124215-2013-07754
Event Type
Injury
Date Received
July 8, 2013
Date of Event
April 29, 2013
Report Date
April 29, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
PMA / PMN Number
P960004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT ATRIAL (RA) LEAD AND RIGHT VENTRICULAR (RV) LEAD WERE FRACTURED DUE TO CLAVICULAR CRUSH. OVERSENSING AND PACING INHIBITION WERE ALSO OBSERVED DUE DAMAGE TO BOTH LEADS. BOTH LEADS WERE SUBSEQUENTLY EXPLANTED AND REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. THE LEADS WERE NOT AVAILABLE FOR RETURN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311195 FINELINE II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4470

Patients

Seq Age Sex Outcome Treatment
1 65 YR Hospitalization| L| R A155| 4470| 0158| E140| 0292