FDA Adverse Event Malfunction Summary report: N

COGNIS

MDR report key: 3210987 · Received July 8, 2013

Report

Report Number
2124215-2013-08145
Event Type
Malfunction
Date Received
July 8, 2013
Date of Event
April 24, 2013
Report Date
April 24, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS SUCCESSFULLY EXPLANTED AND REPLACED. UPON RECEIPT IN OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL INSPECTION CONFIRMED THE CLINICAL OBSERVATION OF A WEAKENED HEADER BOND. IMPROVEMENTS HAVE BEEN MADE TO THE MANUFACTURING PROCESS IN ORDER TO STRENGTHEN THE BOND BETWEEN THE HEADER AND THE DEVICE CASE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING ROUTINE REPLACEMENT OF THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D), THE HEADER WAS OBSERVED TO BE SEPARATED FROM THE DEVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
309889 COGNIS IMPLANTABLE CHF PULSE GENERATOR NIK GUIDANT CRM CLONMEL IRELAND N119

Patients

Seq Age Sex Outcome Treatment
1 76 YR H179| 2188| 4518| N119| 4591| 4087| 4555| 4525| 4524| 4512| 0158