FDA Adverse Event
Malfunction
Summary report: N
COGNIS
MDR report key: 3210987
·
Received July 8, 2013
Report
- Report Number
- 2124215-2013-08145
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Date of Event
- April 24, 2013
- Report Date
- April 24, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NIK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS SUCCESSFULLY EXPLANTED AND REPLACED. UPON RECEIPT IN OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL INSPECTION CONFIRMED THE CLINICAL OBSERVATION OF A WEAKENED HEADER BOND. IMPROVEMENTS HAVE BEEN MADE TO THE MANUFACTURING PROCESS IN ORDER TO STRENGTHEN THE BOND BETWEEN THE HEADER AND THE DEVICE CASE.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING ROUTINE REPLACEMENT OF THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D), THE HEADER WAS OBSERVED TO BE SEPARATED FROM THE DEVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 309889 | COGNIS | IMPLANTABLE CHF PULSE GENERATOR | NIK | GUIDANT CRM CLONMEL IRELAND | N119 |