Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE LEAD IS BEING EVALUATED IN OUR POST MARKET QUALITY ASSURANCE LABORATORY. THIS PRODUCT ISSUE WILL BE UPDATED WHEN ANALYSIS IS COMPLETE.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL INSPECTION OF THE LEAD WAS PERFORMED. INSPECTION OF THE LEAD BODY AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING DID NOT IDENTIFY ANY LEAD CHARACTERISTICS THAT WOULD HAVE RESULTED IN THE CLINICAL OBSERVATION OF DISLODGEMENT. THIS PRODUCT ISSUE WILL BE UPDATED IF ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEAD HAD DISLODGED. A REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS REMOVED FROM SERVICE AND REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.