FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3210959 · Received July 8, 2013

Report

Report Number
2531779-2013-09766
Event Type
Malfunction
Date Received
July 8, 2013
Report Date
June 10, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION FOLLOW-UP #1 SUBMITTED 08/29/2013: THE DEVICE WAS RETURNED TO ANIMAS AND EVALUATED BY PRODUCT ANALYSIS ON 08/08/2013 WITH THE FOLLOWING RESULTS: OBSERVED THE BOLUS BUTTON COVER AND PLUG ARE DETACHED FROM THE PUMP, EXPOSING THE INTERNAL CONTACT. NO EVIDENCE OF CONTAMINATION OR OTHER ANOMALIES FOUND. UNABLE TO OBTAIN AN AUDIBLE ALARM READING BECAUSE OF DAMAGED BOLUS BUTTON.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (ABB TACTILE CNG PRIOR TO DMG) ISSUE. THE REPORTER STATED THAT THE AUDIO BOLUS BUTTON WAS UNRESPONSIVE AND THE AUDIO BOLUS BUTTON WAS PEELING UP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310819 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 12 YR