ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-09774
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Report Date
- June 10, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE TIME AND DATE WAS ACCURATELY SET AT START OF INVESTIGATION. INVESTIGATORS PERFORMED PUMP REWIND, LOAD, AND PRIME WITH NO ALARMS. PUMP WAS EXERCISED FOR 24 HOURS ON A 1 UNIT PER HOUR BASAL RATE WITH NO ALARMS DUPLICATED. EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE PCB BOARD HAD FAILED. THE PUMP WOULD NOT RETAIN THE USER PROGRAMMED DATE AND TIME SETTINGS UPON REMOVAL OF THE PRIMARY AA BATTERY. WHEN A NEW AA BATTERY IS INSERTED THE PUMP DISPLAYS THE DEFAULT DATE AND TIME WHICH MUST BE MANUALLY CONFIRMED (OR RESET) BY THE USER IN ORDER TO PROCEED. PUMP BOOTED TO THE VERIFY SCREEN WITH DIM PINK CONTRAST AND DISPLAY LENS WAS FOUND TO BE SCRATCHED. PUMP COVER WAS REMOVED AND THE OLED SCREEN WAS REMOVED AND REPLACED WITH A NEW TEST SCREEN, CONTRAST RETURNED TO NORMAL WITH TEST SCREEN. VISUAL INSPECTION OF BATTERY COMPARTMENT REVEALED, COMPARTMENT CRACK FROM THREADS TO CASE SEAL. THE OK KEYPAD SYMBOL WAS FOUND TO BE WORN. (B)(6). DEVICE HISTORY RECORD REVIEW WAS CONDUCTED ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE PUMP BOOTED TO THE VERIFY SCREEN WITH DIM PINK CONTRAST. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310507 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR |