FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3210930 · Received July 8, 2013

Report

Report Number
2531779-2013-09809
Event Type
Malfunction
Date Received
July 8, 2013
Report Date
April 26, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DEVICE EVALUATION: ON EXAMINATION, THE KEYPAD WAS NOTED TO BE INTACT. ON TESTING, ALL THE KEYPAD BUTTONS WERE UNRESPONSIVE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION UNDER THE CONTACTS OF ALL THE KEYPAD BUTTONS. THE DISPLAY WAS NOTED TO BE DIM AND DISCOLORED. A TEST DISPLAY WAS ATTACHED TO THE PUMP AND ILLUMINATED PROPERLY WITHOUT DISCOLORATION.

Description of Event or Problem · 1

THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013: ALL KEYPAD BUTTONS WERE UNRESPONSIVE WITH CONTAMINATION UNDER THE CONTACTS OF ALL THE KEYPAD BUTTONS. THE DISPLAY WAS DIM AND DISCOLORED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
309579 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1