ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-09822
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Report Date
- June 10, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS (B)(4) 2013 WITH THE FOLLOWING FINDINGS: DURING TESTING, THE DISPLAY SCREEN WAS FOUND TO BE DIM AND DISCOLORED. A TEST SCREEN WAS INSERTED AND WAS FOUND TO FUNCTION PROPERLY. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A CRACK ALONG THE BATTERY COMPARTMENT EXTENDING FROM THE THREADS TO THE CASE SEAL. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 309571 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR |