FDA Adverse Event Malfunction Summary report: N

ENDOTAK RELIANCE

MDR report key: 3210885 · Received July 8, 2013

Report

Report Number
2124215-2013-07467
Event Type
Malfunction
Date Received
July 8, 2013
Date of Event
November 1, 2012
Report Date
April 24, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PHYSICIAN WILL CONTINUE MONITORING. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A ROUTINE IN-CLINIC FOLLOW UP, REVIEW OF STORED DEVICE MEMORY REVEALED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND IMPLANTABLE DEFIBRILLATION LEAD EXHIBITED A HIGH OUT OF RANGE SHOCK IMPEDANCE MEASUREMENT GREATER THAN 125 OHMS APPROXIMATELY FIVE MONTHS AGO. ALL SUBSEQUENT MEASUREMENTS WERE OBSERVED TO BE WITHIN NORMAL LIMITS. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) DISCUSSED TROUBLESHOOTING OPTIONS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
309227 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI - DEL CARIBE 0180

Patients

Seq Age Sex Outcome Treatment
1 58 YR T180| E143| 4469| 0180