FDA Adverse Event Injury Summary report: N

ALTRUA

MDR report key: 3210869 · Received July 8, 2013

Report

Report Number
2124215-2013-06843
Event Type
Injury
Date Received
July 8, 2013
Date of Event
April 24, 2013
Report Date
June 3, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NVZ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

TO DATE, THE EXPLANTED DEVICE HAS NOT BEEN RECEIVED AT BOSTON SCIENTIFIC. UPON RECEIPT THE DEVICE WILL UNDER GO DETAILED LABORATORY ANALYSIS IN AN ATTEMPT TO CONFIRM AND DETERMINE THE ROOT CAUSE OF THIS EVENT.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS REMOVED AND REPLACED DUE TO SUSPECTED PREMATURE BATTERY DEPLETION (PBD). IN ADDITION THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS. THE DEVICE WAS RETURNED FOR ANALYSIS WHILE THE RIGHT VENTRICULAR (RV) LEAD REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311744 ALTRUA NVZ GUIDANT CRM CLONMEL IRELAND S606

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R