FDA Adverse Event Malfunction Summary report: N

RIATA PASSIVE FIXATION

MDR report key: 3210800 · Received July 8, 2013

Report

Report Number
2938836-2013-03214
Event Type
Malfunction
Date Received
July 8, 2013
Date of Event
May 27, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0457
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING FOLLOW UP, REVIEW OF RECORDS SHOWED SEVERAL EPISODES OF NOISE. EXTERNALIZED CONDUCTORS WERE OBSERVED VIA X-RAY. NO ELECTRICAL ANOMALIES WERE DETECTED. LEAD REMAINS IMPLANTED AND FOLLOW UP HAS BEEN SCHEDULED. IT WAS LATER REPORTED THAT THE PATIENT STOPPED USING HER CLOCK-RADIO AND THE NOISE DISAPPEARED. THE PATIENT CONDITION IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311387 RIATA PASSIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1570/65 NA

Patients

Seq Age Sex Outcome Treatment
1