FDA Adverse Event
Malfunction
Summary report: N
RIATA PASSIVE FIXATION
MDR report key: 3210800
·
Received July 8, 2013
Report
- Report Number
- 2938836-2013-03214
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Date of Event
- May 27, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- Z0457
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING FOLLOW UP, REVIEW OF RECORDS SHOWED SEVERAL EPISODES OF NOISE. EXTERNALIZED CONDUCTORS WERE OBSERVED VIA X-RAY. NO ELECTRICAL ANOMALIES WERE DETECTED. LEAD REMAINS IMPLANTED AND FOLLOW UP HAS BEEN SCHEDULED. IT WAS LATER REPORTED THAT THE PATIENT STOPPED USING HER CLOCK-RADIO AND THE NOISE DISAPPEARED. THE PATIENT CONDITION IS GOOD.