FDA Adverse Event
Injury
Summary report: N
RIATA ST ACTIVE FIXATION
MDR report key: 3210791
·
Received July 8, 2013
Report
- Report Number
- 2938836-2013-03308
- Event Type
- Injury
- Date Received
- July 8, 2013
- Date of Event
- May 28, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWS
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
Additional Manufacturer Narrative · 1
EXTERNAL INSULATION ABRASION WAS NOTED AT 26.0-26.4CM FROM THE CONNECTOR END, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE. THE ETFE COATING WAS INTACT AT THE SAME LOCATION. INSULATION AND CONDUCTOR DAMAGE WAS FOUND AT 24.4-26.0CM FROM THE CONNECTOR END CONSISTENT WITH CLAVICLE CRUSH DAMAGE. THE PACING/SENSING AND SVC CONDUCTORS INSULATION WERE ABRADED. THIS IS CONSISTENT WITH THE OBSERVATION FROM THE FIELD OF NOISE.
Description of Event or Problem · 1
IT WAS REPORTED THAT NOISE WAS OBSERVED ON THE LEAD. DURING EXPLANT, AN INSULATION ANOMALY WAS OBSERVED. LEAD WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 311384 | RIATA ST ACTIVE FIXATION | DEFIBRILLATION LEAD | LWS | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7001/60 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Required Intervention | (B)(4) |