FDA Adverse Event Malfunction Summary report: N

FORTIFY ASSURA VR, DF4 CONNECTOR

MDR report key: 3210660 · Received July 8, 2013

Report

Report Number
2938836-2013-03361
Event Type
Malfunction
Date Received
July 8, 2013
Date of Event
May 13, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
P910023
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHEN AN ASYMPTOMATIC PATIENT PRESENTED IN CLINIC DURING FOLLOW UP, NON SUSTAINED LEAD NOISE EPISODE DUE TO INTERMITTENT UNDERSENSING OF R-WAVES WAS OBSERVED. REPROGRAMMING OF THE DEVICE WAS PERFORMED. ISSUE WAS RESOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
309547 FORTIFY ASSURA VR, DF4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD1257-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 68 YR