UNIFY ASSURA CRT-D

MDR report key: 3210626 · Received July 8, 2013

Report

Report Number
2938836-2013-03462
Event Type
Malfunction
Date Received
July 8, 2013
Date of Event
April 19, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NIK
PMA / PMN Number
P910023
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT MERLIN.NET TRANSMISSION SHOWED NON- SUSTAINED LEAD NOISE DUE TO POST-PACED T-WAVE OVERSENSING WHEN PACING. THE PATIENT DID NOT RECEIVE THERAPY. THE OVERSENSING WAS NOTED ON A STORED EGM. THE PATIENT WAS ASYMPTOMATIC. THE DEVICE WAS REPROGRAMMED SUCCESSFULLY. PATIENT CONDITION IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311725 UNIFY ASSURA CRT-D IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD3257-40 NA

Patients

Seq Age Sex Outcome Treatment
1 51 YR