FDA Adverse Event Malfunction Summary report: N

QUADRA ASSURA CRT-D

MDR report key: 3210573 · Received July 8, 2013

Report

Report Number
2938836-2013-03475
Event Type
Malfunction
Date Received
July 8, 2013
Date of Event
April 22, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NIK
PMA / PMN Number
P910023
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT NON-SUSTAINED LEAD NOISE DUE TO T-WAVE OVERSENSING WAS NOTED VIA A MERLIN.NET TRANSMISSION. DURING A FOLLOW-UP, POST-PACED T-WAVE OVERSENSING ON THE VENTRICULAR LEAD WAS OBSERVED. THE PATIENT DID NOT RECEIVE THERAPY. THE OVERSENSING WAS NOTED ON STORED EGMS. THE PATIENT IS ASYMPTOMATIC. THE DEVICE WAS REPROGRAMMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311522 QUADRA ASSURA CRT-D IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD3265-40 NA

Patients

Seq Age Sex Outcome Treatment
1