RIATA PASSIVE FIXATION
MDR report key: 3210555
·
Received July 8, 2013
Report
- Report Number
- 2938836-2013-03203
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Date of Event
- May 7, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWS
- PMA / PMN Number
- P950022
- Removal / Correction Number
- Z0457
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
NEW INFORMATION NOTES THE LEAD WAS CAPPED DUE TO HIGH VOLTAGE LEAD IMPEDANCE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE ASYMPTOMATIC PATIENT PRESENTED TO CLINIC FOR A FOLLOW UP. EXTERNALIZED CONDUCTORS WERE OBSERVED VIA FLUOROSCOPY. NO ELECTRICAL ANOMALIES WERE NOTED. THE LEAD WILL BE CAPPED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310672 | RIATA PASSIVE FIXATION | DEFIBRILLATION LEAD | LWS | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1570/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR |