FDA Adverse Event Injury Summary report: N

UNIFY QUADRA CRT-D, DF4 CONNECTOR

MDR report key: 3210548 · Received July 8, 2013

Report

Report Number
2938836-2013-03444
Event Type
Injury
Date Received
July 8, 2013
Date of Event
May 10, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NIK
PMA / PMN Number
P030054
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED A PATIENT PRESENTED IN CLINIC AFTER RECEIVING AN INAPPROPRIATE SHOCK. IT WAS DETERMINED INTERMITTENT ATRIAL UNDER-SENSING CAUSED MISDIAGNOSIS WHICH LED TO THERAPY. PROGRAMMING CHANGES WERE RECOMMENDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
309990 UNIFY QUADRA CRT-D, DF4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD3249-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 72 YR