FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR

MDR report key: 3210512 · Received July 8, 2013

Report

Report Number
2938836-2013-03335
Event Type
Injury
Date Received
July 8, 2013
Date of Event
March 29, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE ER AFTER RECEIVING INAPPROPRIATE SHOCKS DUE TO CROSSTALK OF ATRIAL SIGNAL ON THE RIGHT VENTRICULAR CHANNEL. X-RAY REVEALED THE LEAD HAD DISLODGED. CHANGES IN PACING LEAD IMPEDANCE AND THRESHOLD WERE NOTED, BUT STILL WITHIN NORMAL RANGE. LOSS OF CAPTURE WAS ALSO NOTED. THE LEAD WAS EXPLANTED AND REPLACED WHEN REPOSITIONING WAS UNSUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
309860 DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7122Q/58 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention