FDA Adverse Event Injury Summary report: N

RIATA ST ACTIVE FIXATION

MDR report key: 3210502 · Received July 8, 2013

Report

Report Number
2938836-2013-03295
Event Type
Injury
Date Received
July 8, 2013
Date of Event
April 10, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED EXTERNAL INSULATION ABRASION WAS NOTED AT 12.7-12.8CM FROM THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS INTACT AT THE SAME LOCATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAS TWIDDLERS SYNDROME. AN INSULATION BREAK WAS OBSERVED ON THE LEAD. THE LEAD WAS CAPPED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311122 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7000/60 NA

Patients

Seq Age Sex Outcome Treatment
1 61 YR Required Intervention (B)(4)