FDA Adverse Event
Malfunction
Summary report: N
RIATA ACTIVE FIXATION
MDR report key: 3210482
·
Received July 8, 2013
Report
- Report Number
- 2938836-2013-03245
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Date of Event
- May 17, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED INTERNAL INSULATION ABRASION WAS NOTED AT 12.0-12.1 CM AND 30.1-30.2 FROM THE HELIX. THE ETFE COATING WAS INTACT AT THESE LOCATIONS. (B)(6) - NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.