FDA Adverse Event Injury Summary report: N

CURRENT PLUS DR, DF-4 CONNECTOR

MDR report key: 3210441 · Received July 8, 2013

Report

Report Number
2938836-2013-03372
Event Type
Injury
Date Received
July 8, 2013
Date of Event
May 2, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
P910023
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED PREMATURE BATTERY DEPLETION WAS CONFIRMED IN THE LABORATORY. BASED ON THE AVAILABLE PARAMETER AND USAGE INFORMATION, A LONGEVITY CALCULATION WAS PERFORMED AND WAS FOUND TO BE BELOW THE EXPECTED LIMITS. THE ORIGINAL BATTTERY WAS RETURNED TO THE VENDER FOR FURTHER EVALUATION AND NO ANOMALY WAS FOUND. THE CAUSE OF THE PREMATURE BATTERY DEPLETION COULD NOT BE DETERMINED.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE BATTERY VOLTAGE DROPPED SIGNIFICANTLY OVER A PERIOD OF A FEW DAYS AND REACHED ERI. THE PATIENT WAS NOT PACEMAKER DEPENDENT. THE DEVICE WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
309731 CURRENT PLUS DR, DF-4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD2211-36Q NA

Patients

Seq Age Sex Outcome Treatment
1 57 YR Required Intervention