FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR

MDR report key: 3210437 · Received July 8, 2013

Report

Report Number
2938836-2013-03339
Event Type
Injury
Date Received
July 8, 2013
Date of Event
April 18, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DUE TO LEAD DISLODGEMENT, THE DEVICE INTERMITTENTLY UNDERSENSED VENTRICULAR FIBRILLATION AND FAILED TO CONVERT THE PATIENT. CARDIAC RESUSCITIATION WAS PERFORMED. ONCE THE PATIENT WAS STABILIZED, THE LEAD WAS REPOSITIONED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311397 DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7122Q/58 NA

Patients

Seq Age Sex Outcome Treatment
1 78 YR Required Intervention