FDA Adverse Event Injury Summary report: N

RIATA PASSIVE FIXATION

MDR report key: 3210409 · Received July 8, 2013

Report

Report Number
2938836-2013-03218
Event Type
Injury
Date Received
July 8, 2013
Date of Event
April 25, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT T-WAVE OVERSENSING WAS OBSERVED. FOUR DAYS LATER, DFT TESTING WAS PERFORMED TO TEST THE DECREASED SETTINGS WHEN UNDERSENSING OF VF WAS OBSERVED. LEAD WAS CAPPED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310657 RIATA PASSIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1571/65 NA

Patients

Seq Age Sex Outcome Treatment
1 75 YR Required Intervention CD3231-40, 739362