FDA Adverse Event
Injury
Summary report: N
EASYTRAK 3
MDR report key: 3210365
·
Received July 8, 2013
Report
- Report Number
- 2124215-2013-07729
- Event Type
- Injury
- Date Received
- July 8, 2013
- Date of Event
- April 22, 2013
- Report Date
- June 10, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- OJX
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL INSPECTION OF THE LEAD WAS PERFORMED. INSPECTION OF THE LEAD BODY AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING DID NOT IDENTIFY ANY LEAD CHARACTERISTICS THAT WOULD HAVE RESULTED IN THE CLINICAL OBSERVATION OF DISLODGEMENT.
Additional Manufacturer Narrative · 1
THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
Description of Event or Problem · 1
--
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD WAS FOUND TO BE DISLODGED INTO THE ATRIUM. THE LEAD WAS EXPLANTED AND REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310556 | EASYTRAK 3 | IMPLANTABLE LEAD | OJX | CPI - DEL CARIBE | 4548 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| L| R |