FDA Adverse Event Injury Summary report: N

VITALITY

MDR report key: 3210346 · Received July 8, 2013

Report

Report Number
2124215-2013-07709
Event Type
Injury
Date Received
July 8, 2013
Date of Event
April 17, 2013
Report Date
May 24, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS PRODUCT IS CURRENTLY UNDERGOING DETAILED ANALYSIS. THIS EVENT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED MULTIPLE SHOCKS. A REVIEW OF THE ELECTROGRAM (EGM) SHOWED NOISE THAT WAS OVERSENSED BY THE RIGHT VENTRICULAR (RV) LEAD. ISOMETRICS WERE PERFORMED BUT THE NOISE COULD NOT BE REPRODUCED. THE DEVICE WAS PROGRAMMED TO MONITOR ONLY TO KEEP THE PATIENT FROM RECEIVING MORE SHOCKS. A PROCEDURE WAS PERFORMED, THE DEVICE WAS EXPLANTED AND THE RV LEAD WAS SURGICALLY ABANDONED. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
309603 VITALITY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T125

Patients

Seq Age Sex Outcome Treatment
1 61 YR Hospitalization| L| R 4068| 1861| 0155| T125