FDA Adverse Event Malfunction Summary report: N

ENDOTAK DSP

MDR report key: 3210298 · Received July 8, 2013

Report

Report Number
2124215-2013-07548
Event Type
Malfunction
Date Received
July 8, 2013
Date of Event
January 1, 2012
Report Date
July 13, 2016
Manufacturer
CPI - DEL CARIBE
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH OUT-OF-RANGE (OOR) PACING IMPEDANCES OF GREATER THAN 2000 OHMS, AS WELL AS OOR SHOCKING IMPEDANCES OF GREATER THAN 200 OHMS. ADDITIONALLY THERE WAS NOISE NOTED ON THE LEAD WHICH LEAD TO NONSUSTAINED EVENTS. AS THE PATIENT HAS NO INSURANCE THEY ARE REFUSING TREATMENT. THE DEVICE WAS ELECTRICALLY TURNED OFF TO AVOID ANY ADDITIONAL ISSUES WITH THE LEAD AND THE PATIENT IS AWARE OF THE COMPLICATIONS THAT COULD ARISE SHOULD THEY NEED THERAPY WHILE THE DEVICE IS PROGRAMMED OFF. TO DATE, NO ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION RECEIVED INDICATES THAT THE PATIENT WAS IN A CAR ACCIDENT WHICH DAMAGED HIS SHOULDER AND THE RIGHT VENTRICULAR (RV) LEAD. THIS OCCURRED A YEAR PREVIOUS TO THE OBSERVED OUT-OF-RANGE PACE IMPEDANCE MEASUREMENTS. THIS PRODUCT REMAINS IMPLANTED BUT REMAINS PROGRAMMED OFF. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
309333 ENDOTAK DSP IMPLANTABLE LEAD LWS CPI - DEL CARIBE 0095

Patients

Seq Age Sex Outcome Treatment
1 58 YR E102| 1600| 0095| 1743| 1860| MISMATCH