PEDIATRIC DISTRACTOR ASSEMBLY
Report
- Report Number
- 1526439-2013-20633
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Report Date
- June 17, 2013
- Manufacturer
- DEPUY SYNTHES SPINE
- Product Code
- LXH
- PMA / PMN Number
- PEXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- PHYSICIAN
Narratives
A FOLLOW UP REPORT WILL BE FILED UPON RECEIPT OF THE DEVICE AND COMPLETION OF THE INVESTIGATION.
VISUAL INSPECTION OF THE RETUNED PEDIATRIC DISTRACTOR ASSEMBLY [PRODUCT CODE: 6983-27-722, LOT NO: JC786201] NOTED THAT ONE OF THE DISTIL TIP¿S FEET HAD BECOME BROKEN. THE BROKEN DISTAL TIP¿S FOOT WAS NOT RETURNED FOR EVALUATION AS IT WAS DISCARDED. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR THE PEDIATRIC DISTRACTOR ASSEMBLY [PRODUCT CODE: 6983-27-722, LOT NO: JC786201] WAS CONDUCTED IDENTIFYING THAT A LOT QUANTITY OF 1 UNIT WAS RELEASED TO STOCK ON 5/8/12 WITH NO DISCREPANCIES. NO ISSUES WERE IDENTIFIED DURING THE MANUFACTURING AND RELEASE OF THIS PRODUCT THAT COULD HAVE BEEN CONTRIBUTED TO THE PROBLEM REPORTED BY THE CUSTOMER. THE PRODUCT WAS RELEASED ACCOMPLISHING ALL QUALITY REQUIREMENTS. A REVIEW OF THE COMPLAINT TREND ANALYSIS FOUND NO OBSERVED TRENDS FOR ISSUES OF THIS NATURE. THE ROOT CAUSE OF THE FRACTURED DISTAL TIP¿S FOOT CANNOT POSITIVELY BE DETERMINED. HOWEVER, THE NOTED DAMAGE ON THE DISTAL TIP¿S FOOT SUGGESTS THAT IT HAS BEEN SUBJECTED TO AN UNANTICIPATED AXIAL STRESS RESULTING IN A FRACTURE. AS THERE HAS BEEN NO ISSUE IDENTIFIED IN THE MANUFACTURING OR RELEASE OF THE DEVICE, AND THERE HAS BEEN NO SYSTEMATIC TREND, THIS FILE WILL BE CLOSED WITH NO FURTHER ACTION REQUIRED.
IT WAS REPORTED THAT THE PIN ON THE DISTAL END OF THE DISTRACTOR ASSEMBLY BROKE OFF AND THE INSTRUMENT BECAME BENT DURING USAGE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT AND NO SIGNIFICANT DELAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310047 | PEDIATRIC DISTRACTOR ASSEMBLY | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | LXH | DEPUY SYNTHES SPINE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |