ALTRUA
Report
- Report Number
- 2124215-2013-08499
- Event Type
- Injury
- Date Received
- July 8, 2013
- Date of Event
- April 19, 2013
- Report Date
- May 31, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NVZ
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- PHYSICIAN
Narratives
THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.
--
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT IMPLANTED WITH THIS PACEMAKER AND RIGHT VENTRICULAR (RV) LEAD WAS SYMPTOMATIC DUE TO INTERMITTENT EXIT BLOCK. THE CAUSE OF THE EXIT BLOCK WAS UNABLE TO BE DETERMINED. PACING INHIBITION WAS OBSERVED AND AN INCREASE IN PACING THRESHOLD MEASUREMENTS WAS NOTED. AN INVASIVE PROCEDURE WAS PERFORMED. THIS DEVICE WAS SUCCESSFULLY EXPLANTED AND THE RV LEAD WAS SURGICALLY ABANDONED. A NEW SYSTEM WAS IMPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310003 | ALTRUA | IMPLANTABLE PULSE GENERATOR | NVZ | GUIDANT CRM CLONMEL IRELAND | S603 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Hospitalization| L| R | 4469| 1290| 4087| S603 |