FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3210188 · Received July 3, 2013

Report

Report Number
8020893-2013-01500
Event Type
Malfunction
Date Received
July 3, 2013
Date of Event
January 1, 2013
Report Date
June 6, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) REPORTED THAT AN 840 VENTILATOR HAD AN ERRATIC DISPLAY. NO PATIENT INVOLVEMENT. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHIC USER INTERFACE (GUI) CPU PCB AND LCD DISPLAY PANEL. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
304634 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1