FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3210168 · Received July 3, 2013

Report

Report Number
8020893-2013-01518
Event Type
Malfunction
Date Received
July 3, 2013
Date of Event
May 5, 2013
Report Date
June 7, 2013
Manufacturer
COVIDIEN
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT AN 840 VENTILATOR HAD A BLANK SCREEN. THERE WAS NO PATIENT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GUI COMMUNICATIONS LCD (LIQUID CRYSTAL DISPLAY) PANEL. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
305999 840 VENTILATOR VENTILATOR CBK COVIDIEN 840

Patients

Seq Age Sex Outcome Treatment
1