FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3210098 · Received July 3, 2013

Report

Report Number
1720753-2013-07850
Event Type
Malfunction
Date Received
July 3, 2013
Date of Event
June 26, 2013
Report Date
July 3, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED AN ONSITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE IDENTIFIED OR DUPLICATED. THE SYS WAS OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A BLACK SCREEN WHEN ATTEMPTING TO TAKE A FLUORO SHOT. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
305656 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1