FDA Adverse Event Injury Summary report: N

ACUITY

MDR report key: 3210088 · Received July 8, 2013

Report

Report Number
2124215-2013-07391
Event Type
Injury
Date Received
July 8, 2013
Date of Event
April 13, 2013
Report Date
May 15, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD EXHIBITED LOSS OF CAPTURE DUE TO DISLODGEMENT. AN INVASIVE PROCEDURE WAS PERFORMED AND THIS LEAD WAS SUCCESSFULLY REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310373 ACUITY IMPLANTABLE HF LEAD OJX CPI - DEL CARIBE 4592

Patients

Seq Age Sex Outcome Treatment
1 69 YR Hospitalization| L| R 4592| N160| 0295| 4470| 4543| 4555