FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3210052 · Received June 27, 2013

Report

Report Number
3008642652-2013-01704
Event Type
Malfunction
Date Received
June 27, 2013
Date of Event
June 1, 2013
Report Date
June 25, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (ADJUST BELT, CHECK THERAPY ELECTRODE ALARMS) HAS BEEN CONFIRMED. UPON INVESTIGATION THE WHITE PULSE WIRE WAS FOUND TO BE OPEN IN THE TRUNK CABLE INSULATION, CAUSING THE REPORTED ALARMS. THE ROOT CAUSE FOR THE OPEN PULSE WIRE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE ON THE TRUNK CABLE SECTION. NO ADVERSE EVENT RESULTED FROM THE OPEN PULSE WIRE. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A ZOLL PATIENT SERVICE REPRESENTATIVE (PSR) CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT A (B)(6) MALE PATIENT WAS RECEIVING ADJUST BELT AND CHECK THERAPY ELECTRODE ALARMS. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
292329 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 56 YR