ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2013-07382
- Event Type
- Injury
- Date Received
- July 8, 2013
- Date of Event
- April 17, 2013
- Report Date
- October 8, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- LWS
- PMA / PMN Number
- P910073
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE ICD PRODUCT WAS RETURNED FOR ANALYSIS; THE LEAD WAS SURGICALLY ABANDONED. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ICD ANALYSIS.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. LABORATORY ANALYSIS DID NOT IDENTIFY ANY DEVICE CHARACTERISTICS THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED CLINICAL OBSERVATIONS.
--
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED IN THE NEAR FUTURE, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
--
--
APPROXIMATELY ONE MONTH LATER, THE PATIENT RETURNED FOR EVALUATION AND REVISION. NOISE WAS REPRODUCIBLE BEFORE AND DURING THE PROCEDURE WITH MANEUVERS WHILE CONFIRMING CONNECTIONS. NO NOTICEABLE LEAD DAMAGE WAS OBSERVED. X-RAY ANALYSIS REVEALED INSERTION ISSUES WITH THIS LEAD'S TERMINAL PINS. THE PHYSICIAN ELECTED TO REPLACE THE ICD AND SURGICALLY ABANDONED THE LEAD. A NEW NON-BOSTON SCIENTIFIC DEVICE AND BOSTON SCIENTIFIC LEAD COMBINATION WAS SUCCESSFULLY IMPLANTED. THE ICD WAS FOR ANALYSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS OCCURRED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT RECENT NOISE EPISODES STORED ON THIS PATIENT SYSTEM ARE MOST LIKELY RELATED TO A LEAD CONNECTION OR LEAD INTEGRITY ISSUE. ADDITIONALLY, THE PATIENT RECEIVED INAPPROPRIATE SHOCKS AS A RESULT OF RHYTHM ID DEFAULT. MUCH INVESTIGATION HAS PURSUED THE PAST WEEK WITH BOSTON SCIENTIFIC TECHNICAL THERAPY SUPPORT (TTS) EVALUATING PRINTOUTS AND DATA DISK ANALYSIS OF STORED EPISODES. IT IS UNDERSTOOD THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (F102; ICD) AND THE RIGHT VENTRICULAR (0181; RV) LEAD WERE IMPLANTED ABOUT TWO YEARS AGO. APPROXIMATELY TWO MONTHS POST-IMPLANT, NOISE WAS RECORDED ON THE SHOCK CHANNEL, AS EVIDENT BY REVIEW OF HISTORICAL EPISODES. TTS CONFIRMED THAT THE RECORDED NOISE IS CONSISTENT WITH A LEAD OR WITH A CONNECTION ISSUE. THE NOISE IS LOW FREQUENCY, RANDOMLY DISTRIBUTED WITH HIGH AMPLITUDE. ADDITIONALLY, FROM TREND HISTORICAL TREND GRAPH REVIEW, BOTH THE RV AND SHOCK IMPEDANCES FLUCTUATE BUT THE GENERAL TREND IS QUITE STABLE WITH A DECREASE BEGINNING EARLIER THIS MONTH. ADDITIONAL TROUBLESHOOTING WAS PERFORMED CLINICALLY, AND THE NOISE WAS REPRODUCIBLE WITH ARM STRETCHING AND OTHER RELATED MANOEUVRES. TTS RECOMMENDED AN INTENSIVE INVESTIGATION ON THIS LEAD, AND A SYSTEM INTEGRITY TEST TO DETERMINE THE ROOT CAUSE OF THIS ISSUE. THE PHYSICIAN PLANS TO SEE THE PATIENT IN SHORTENED INTERVALS, AS WELL AS TO PROVIDE REMOTE MONITORING. THERE HAVE BEEN NO ADVERSE PATIENT EFFECTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310099 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | LWS | CPI - DEL CARIBE | 0181 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| L| R |