FDA Adverse Event Injury Summary report: N

TELIGEN

MDR report key: 3210036 · Received July 8, 2013

Report

Report Number
2124215-2013-07293
Event Type
Injury
Date Received
July 8, 2013
Date of Event
April 17, 2013
Report Date
May 7, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE RV LEAD WAS SURGICALLY ABANDONED AND REPLACED AND THE DEVICE WAS ELECTIVELY EXPLANTED AND REPLACED AS IT HAD BEEN IMPLANTED SUB-PECTORAL. THIS DEVICE IS INCLUDED IN THE (B)(6) 2009 SUBPECTORAL IMPLANT 2009 ADVISORY POPULATION. THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. LABORATORY ANALYSIS DID NOT IDENTIFY ANY DEVICE CHARACTERISTICS THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED CLINICAL OBSERVATIONS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) IN ASSOCIATION WITH ANOTHER MANUFACTURER'S RIGHT VENTRICULAR (RV) LEAD, EXHIBITED NOISE. THE NOISE WAS OVERSENSED RESULTING IN AN UNKNOWN DURATION OF PACING INHIBITION. AN INVASIVE PROCEDURE WAS PERFORMED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311733 TELIGEN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND E110

Patients

Seq Age Sex Outcome Treatment
1 46 YR Hospitalization| L| R MISMATCH| 7120| 0293| E142| E110| 1888TC