FDA Adverse Event Malfunction Summary report: N

COULTER® HMX AUTOLOADER ANALYZER (115V)

MDR report key: 3207942 · Received July 5, 2013

Report

Report Number
1061932-2013-01378
Event Type
Malfunction
Date Received
July 5, 2013
Date of Event
June 7, 2013
Report Date
June 7, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K010765
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT. THE FSE FOUND THAT THE TUBING THROUGH PINCH VALVE PV43 WAS WORN. THE FSE REPLACED THE TUBING AND THE LEAK WAS FIXED. THE REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES. PER LABELING, BECKMAN COULTER URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER (BEC) REPORTING A LEAK INSIDE THE COULTER HMX AUTOLOADER ANALYZER (115V). THE VOLUME OF THE LEAK WAS ABOUT 20 ML AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF A LABORATORY COAT, GLOVES, AND PROTECTIVE EYEWEAR AT THE TIME OF THE OCCURRENCE. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED. THE LABORATORY'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. MEDICAL ATTENTION WAS NOT SOUGHT. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION TO THIS EVENT.