COULTER® LH 750 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2013-01377
- Event Type
- Malfunction
- Date Received
- July 5, 2013
- Date of Event
- June 7, 2013
- Report Date
- June 7, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K061574
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT. THE FSE FOUND A DEFECTIVE ASPIRATION NEEDLE. THE FSE REPLACED THE ASPIRATION NEEDLE WHICH RESOLVED THE LEAK. PER LABELING, BECKMAN COULTER URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IS NOT LIMITED TO PROTECTIVE EYEWEAR, GLOVES AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).
THE CUSTOMER CONTACTED BECKMAN COULTER (BEC) STATING THAT THERE WAS A LEAK OF CLEAR CLENZ ON THE LEFT SIDE OF THE COULTER LH 750 HEMATOLOGY ANALYZER. THE VOLUME OF THE LEAK WAS ABOUT 5 MLS AND WAS NOT CONTAINED WITHIN THE UNIT. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND PROTECTIVE EYEWEAR AT THE TIME OF THE EVENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS REVIEWED. THE LABORATORY'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION TO THE REPORTED EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 307685 | COULTER® LH 750 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | LH 750 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |