FDA Adverse Event Malfunction Summary report: N

COULTER® LH 750 HEMATOLOGY ANALYZER

MDR report key: 3207938 · Received July 5, 2013

Report

Report Number
1061932-2013-01377
Event Type
Malfunction
Date Received
July 5, 2013
Date of Event
June 7, 2013
Report Date
June 7, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061574
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT. THE FSE FOUND A DEFECTIVE ASPIRATION NEEDLE. THE FSE REPLACED THE ASPIRATION NEEDLE WHICH RESOLVED THE LEAK. PER LABELING, BECKMAN COULTER URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IS NOT LIMITED TO PROTECTIVE EYEWEAR, GLOVES AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BECKMAN COULTER (BEC) STATING THAT THERE WAS A LEAK OF CLEAR CLENZ ON THE LEFT SIDE OF THE COULTER LH 750 HEMATOLOGY ANALYZER. THE VOLUME OF THE LEAK WAS ABOUT 5 MLS AND WAS NOT CONTAINED WITHIN THE UNIT. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND PROTECTIVE EYEWEAR AT THE TIME OF THE EVENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS REVIEWED. THE LABORATORY'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION TO THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
307685 COULTER® LH 750 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER LH 750 NA

Patients

Seq Age Sex Outcome Treatment
1