FDA Adverse Event Other Summary report: N

TANGORS

MDR report key: 3205157 · Received June 28, 2013

Report

Report Number
9612420-2013-00007
Event Type
Other
Date Received
June 28, 2013
Date of Event
May 29, 2013
Report Date
June 28, 2013
Manufacturer
ULRICH GMBH & CO. KG
Product Code
NKB
PMA / PMN Number
K052385
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

ULRICH MEDICAL HAS IDENTIFIED A LABELING ISSUE ON PART NUMBER CS 3008-045, 6.0 MM DIAMETER - 45 MM TITANIUM ROD FOR THE TANGO RS SCREW AND ROD SYSTEM. SEVERAL LOTS WERE LASER MARKED AS CS 3008-040 (6.0 MM DIAMETER, 40 MM LENGTH) ON THE ROD WITH A LENGTH OF 45 MM. THERE IS NO RISK TO THE PATIENT, AS THE SURGEON WILL MEASURE THE NEEDED LENGTH OF THE ROD INSITU WITH A SEPARATE INSTRUMENT. THE INSTRUMENT IS THEN TRANSFERRED TO THE TRAY TO DETERMINE THE APPROPRIATE LENGTH OF THE ROD. LASER MAKING PART NUMBERS ON THE ROD ARE NOT CONFIRMED BY THE SURGEON BUT THE LENGTH OF THE ROD PER THE MEASURING TOOL. THE ISSUE WAS IDENTIFIED DURING INCOMING GOODS INSPECTION BY ONE OF ULRICH MEDICAL'S CUSTOMERS. (B)(4). ULRICH MEDICAL WILL INFORM ALL OTHER INTERNATIONAL CUSTOMERS OF AFFECTED LOTS AND HAVE THEM IMMEDIATELY QUARANTINED. UNITS WILL BE SHIPPED BACK TO ULRICH MEDICAL AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
295835 TANGORS TANGORS ROD, BLUE, 0 6, 0 MM, LENGTH 45 MM NKB ULRICH GMBH & CO. KG CS 3008-045 WSA14719871

Patients

Seq Age Sex Outcome Treatment
1 NA