FDA Adverse Event Malfunction Summary report: N

MICRO EYE SCOPE

MDR report key: 32038 · Received February 22, 1996

Report

Report Number
32038
Event Type
Malfunction
Date Received
February 22, 1996
Date of Event
August 11, 1995
Report Date
November 30, 1995
Manufacturer
CARL ZEISS, INC.
Product Code
EPT
Product Problem
Yes
Report Source
User Facility report
Reporter Location
LA, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

DURING (RIGHT EYE) CATARACT PHACOEMULSIFICATION WITH LENS IMPLANTATION AND WHILE LOWERING THE MICROSCOPE INTO POSITION BY THE SURGEON, THE SCOPE CAME LOOSE AND STARTED TO FALL ONTO PT. THE CIRCULATING OR NURSE GRABBED UNDER AND HELD THE SCOPE. PHYSICIAN AND NURSE PUT SCOPE BACK INTO POSITION AND REATTACHED IT. CO TECH FOUND THAT THE SECURITY SCREW COULD HAVE WORKED ITS WAY OUT DUE TO BREAKDOWN IN LUBRICATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICRO EYE SCOPE MICRO EYE SCOPE EPT CARL ZEISS, INC.

Patients

Seq Age Sex Outcome Treatment
1 57 YR