PRECISE BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2013-02408
- Event Type
- Malfunction
- Date Received
- July 3, 2013
- Date of Event
- May 27, 2013
- Report Date
- June 6, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED THE REPORTED COMPLAINT. FOR CLARIFICATION, THE PITCH CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. ADDITIONAL OBSERVATIONS NOT REPORTED BY THE SITE WERE A CONTAMINATED DALLAS CHIP. AFTER REMOVING THE HOUSING, A DARK LAYER OF RESIDUE WAS FOUND ON THE DALLAS CHIP. ENGINEERING CONCLUDED THE DAMAGE WAS LIKELY CLEANING RELATED. ELECTRICAL CONTINUITY PASSED. NO OTHER DAMAGE WAS FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: WARNING: READ ALL INSTRUCTIONS CAREFULLY. FAILURE TO PROPERLY FOLLOW INSTRUCTIONS MAY CAUSE IMPROPER FUNCTIONING OF THE DEVICE. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
THE INSTRUMENT HAS NOT BEEN RETURNED FOR EVALUATION, THEREFORE, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE INSTRUMENT IS RETURNED (POST ENGINEERING EVALUATION) OR IF ADDITIONAL INFORMATION IS RECEIVED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI S PROSTATECTOMY PROCEDURE A TORN STRING WAS DETECTED. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 305210 | PRECISE BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420110-06 | M10120522 282 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS & ACCESSORIES ESU| DA VINCI S SYSTEM INSTRUMENTS & ACCESSORIES ESU |