FDA Adverse Event Malfunction Summary report: N

VENTED AUTOFEED CHAMBER

MDR report key: 3203620 · Received July 3, 2013

Report

Report Number
9611451-2013-00505
Event Type
Malfunction
Date Received
July 3, 2013
Date of Event
June 3, 2013
Report Date
June 3, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290V CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) WHERE IT WAS CONNECTED TO A WATER BAG TO TEST FOR THE REPORTED LEAK. RESULTS: WHEN THE RETURNED WATER FEEDSET TUBE WAS CONNECTED TO A WATER BAG AND PLACED UNDER TENSION A SMALL DROP OF WATER BEGAN TO BUILD AT THE FEEDSET TUBE AND WATERBAG SPIKE CONNECTION. A SUFFICIENT AMOUNT OF GLUE WAS IDENTIFIED AT THE SPIKE TUBING CONNECTION. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 120614. CONCLUSION: THE REPORTED WATER LEAK WAS MOST LIKELY DUE TO THE WATER FEEDSET TUBE BEING SET UP UNDER TENSION. THE SPECIFICATION FOR THE CHAMBER REQUIRES THAT THE FEEDSET TUBE SHOULD HAVE A BREAKING STRAIN OF 30 NEWTONS. DURING PRODUCTION, PULL TESTING OF THE FEEDSET STRENGTH AT BOTH SPIKE AND DOME END IS PERFORMED EVERY HOUR ON MR290 CHAMBERS FROM EACH PRODUCTION LINE. TESTING INVOLVING SIMULATED AGEING OF THE FEEDSET TO CHECK WHETHER THE STRENGTH OF THE GLUE JOINT DECREASES OVER TIME REVEALED THAT THE FEEDSETS ON AGED CHAMBERS WERE ONLY BREAKING AT 50 TO 55N OR MORE, PROVING THAT SHELF LIFE HAS NO NEGATIVE IMPACT ON THE ULTIMATE TENSILE STRENGTH OF THE GLUE JOINT. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. THIS SUGGESTS THAT THE RETURNED FEEDSET TUBE ONLY STARTED TO LEAK AFTER IT WAS RELEASED FOR DISTRIBUTION. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 CHAMBER STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARMS." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT."

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE THAT WATER WAS LEAKING FROM THE CONNECTION BETWEEN THE BAG SPIKE AND WATER FEEDSET TUBE ON AN MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER. THIS WAS FOUND AFTER ONE DAY OF USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
306583 VENTED AUTOFEED CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 120614

Patients

Seq Age Sex Outcome Treatment
1 RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT