FDA Adverse Event Injury Summary report: N

SYNCHROMED II

MDR report key: 3203579 · Received July 3, 2013

Report

Report Number
3004209178-2013-04798
Event Type
Injury
Date Received
July 3, 2013
Report Date
August 22, 2012
Manufacturer
MDT PUERTO RICO OPERATIONS CO
Product Code
LKK
PMA / PMN Number
P860004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT ID: 8835 LOT# SERIAL# UNK, PRODUCT TYPE PROGRAMMER, PATIENT PRODUCT ID: 8731SC LOT# SERIAL# (B)(4), IMPLANTED: 2009 (B)(6), PRODUCT TYPE CATHETER. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT FELL. FOLLOWING THE FALL, THE PATIENT WAS HOSPITALIZED FOR A WEEK, AND AT THE TIME OF THE REPORT WAS IN A REHAB CENTER. SINCE THE PATIENT FELL, THE PERSONAL THERAPY MANAGER (PTM) WAS HAVING DIFFICULTY COMMUNICATING WITH THE PUMP. THEY WERE GETTING THE "COMMUNICATION UNSUCCESSFUL" SCREEN. IT WAS TAKING ABOUT 10 ATTEMPTS TO COMMUNICATE AND GET THE BOLUS. IT WAS NOTED THAT PATIENT USED THE PTM ONLY WHEN NEEDED, WHICH WASN'T VERY FREQUENT. THERE WAS CONCERN THAT THE PUMP MAY HAVE TURNED OR FLIPPED DUE TO THE FALL, AND WONDERED IF THIS COULD BE CONTRIBUTING TO THE COMMUNICATION ISSUES. THE PUMP CONTAINED A MIXTURE OF MORPHINE AND CLONIDINE. ABOUT 1 ½ MONTHS LATER IT WAS REPORTED THAT THE PUMP WAS REFILLED (B)(6) 2012. FOLLOWING THE REFILL WHEN THE PATIENT FIRST ATTEMPTED TO USE THE PTM AN 8503 BOLUS IN PROGRESS WAS NOTED. IT WAS ALSO NOTED THAT THE REFILL DATE HADN¿T CHANGED. THE PATIENT WAS DIRECTED TO THE PHYSICIAN TO DECOUPLE AND RECOUPLE THE PTM TO CORRECT THE REFILL DATE ON THE PTM. IT WAS NOTED THAT AS OF THE DATE OF THE ADDITIONAL INFORMATION BACLOFEN HAD BEEN ADDED TO THE PUMP. CONTINUED ISSUES WITH COUPLING WERE ALSO STILL NOTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
306292 SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE LKK MDT PUERTO RICO OPERATIONS CO 8637-20

Patients

Seq Age Sex Outcome Treatment
1 00068 YR Hospitalization