STERRAD 100NX STERILIZER
Report
- Report Number
- 2084725-2013-00305
- Event Type
- Malfunction
- Date Received
- July 2, 2013
- Date of Event
- June 20, 2013
- Report Date
- June 24, 2013
- Manufacturer
- ADVANCED STERILIZATION PRODUCTS
- Product Code
- MLR
- PMA / PMN Number
- K071385
- Removal / Correction Number
- Z-1026-1027-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER WAS DISPATCHED TO CUSTOMER SITE. ODOR/SMELL WAS CONFIRMED. THE VACUUM PUMP OIL, CATALYTIC DECOMP FILTER AND OIL MIST FILTER WERE REPLACED. UNIT MEETS SPECIFICATIONS AND WAS RETURNED TO SERVICE.
CORRECTION: SEX IS UNKNOWN. MANUFACTURER DATE: 05/26/2011. CONCLUSION: ROOT CAUSE IDENTIFIED IN CAPA WAS FOUND TO BE PREMATURE FAILURE/SATURATED OIL MIST FILTER OR VACUUM PUMP OIL. ASP INVESTIGATION SUMMARY: THE INVESTIGATION INCLUDED A REVIEW OF THE DEVICE HISTORY RECORD, SERVICE HISTORY, TRENDING OF THE PRODUCT MALFUNCTION CODE, FAILURE MODE AND EFFECTS ANALYSIS, HEALTH HAZARD EVALUATION, SYSTEM HAZARD AND USER MISUSE ANALYSIS, AND CAPA. THE DHR (DEVICE HISTORY RECORD) WAS REVIEWED AND NO ISSUES RELATING TO THE FAILURE MODE WERE NOTED. THE INVOLVED UNIT MET MANUFACTURER SPECIFICATIONS AT THE TIME OF RELEASE. THE SERVICE HISTORY FOR THIS UNIT FOR THE PAST 6 MONTHS 12/25/2012 TO 06/23/2013) DID NOT REVEAL A SIGNIFICANT TREND FOR THIS SAME ISSUE. THE TREND OF THE PRODUCT MALFUNCTION CODE ODOR/SMELLS WAS COMPLETED FROM 01/2013 THROUGH 12/2013 AND REVEALED A SIGNIFICANT TREND WHICH WAS ADDRESSED THROUGH CAPA. THE FMEA (FAILURE MODE AND EFFECTS ANALYSIS) REVEALED THE RISK PRIORITY NUMBER SCORES ARE BELOW 100 AND ARE CONSIDERED ACCEPTABLE. THE HHE (HEALTH HAZARD EVALUATION) WAS REVIEWED FOR THE RISK OF ODOR AND SMELL EXPOSURE. THE SEVERITY AND OCCURRENCE FOR THE GENERAL POPULATION WERE LIMITED (TRANSIENT, MINOR IMPAIRMENT, NO MEDICAL TREATMENT REQUIRED) AND THE PRODUCT PROBLEM HAS BEEN KNOWN TO RESULT IN THE IDENTIFIED HARM, BUT ONLY OCCASIONALLY AND/OR UNDER UNUSUAL CIRCUMSTANCES. THE SHUMA (SYSTEM HAZARD AND USER MISUSE ANALYSIS) SHOWS THAT THE RISK IS AS LOW AS REASONABLY PRACTICAL FOR EXPOSURE TO ODOR OR ODORANTS. THE CAPA (CORRECTIVE AND PREVENTATIVE ACTION) IDENTIFIED THE ROOT CAUSE FOR THE ODOR/SMELLS ISSUE AS: (1) PREMATURE FAILURE OF THE USED AND SATURATED STERRAD® 100NX OIL MIST FILTER CAUSED OIL VAPOR EMISSIONS THAT EXACERBATED THE ODOR/SMELL COMPLAINTS REPORTED FOR THE STERRAD® 100NX SYSTEM. (2) THE USE OF A LESS OXIDATIVELY STABLE VACUUM PUMP OIL CAUSED THE ODOR/SMELLS FOR THE STERRAD® 100NX SYSTEM. NO PARTS WERE RETURNED FOR FURTHER EVALUATION. THE ISSUE HAS BEEN RESOLVED AT THE CUSTOMER FACILITY. ASP WILL CONTINUE TO TRACK AND TREND THIS ISSUE.
AN INTERNATIONAL CUSTOMER REPORTED AN EVENT OF A "BAD SMELL" (ODOR) EMITTING FROM THE STERRAD 100NX STERILIZER. THERE WAS NO REPORT OF ANY HUMAN REACTION. A FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY EVENT DATED (B)(6) 2013.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 302892 | STERRAD 100NX STERILIZER | STERRAD EQUIPMENT (MLR) | MLR | ADVANCED STERILIZATION PRODUCTS | NA | 110170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |