STERRAD 50 STERILIZATION SYSTEM
Report
- Report Number
- 2084725-2013-00297
- Event Type
- Malfunction
- Date Received
- July 2, 2013
- Date of Event
- June 27, 2013
- Report Date
- June 27, 2013
- Manufacturer
- ADVANCED STERILIZATION PRODUCTS
- Product Code
- MLR
- PMA / PMN Number
- K981625
- Removal / Correction Number
- Z-0744-0746-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
NI
BASED ON CAPA INVESTIGATION AND ASTM TESTING, OIL DEGRADATION CAN CAUSE ODOR/SMELLS FOR STERRAD SYSTEMS. ASP INVESTIGATION SUMMARY: THE INVESTIGATION INCLUDED A REVIEW OF THE DEVICE HISTORY RECORD (DHR), SERVICE HISTORY, TRENDING OF THE PRODUCT MALFUNCTION CODE, FAILURE MODE AND EFFECTS ANALYSIS (FMEA), SYSTEM HAZARD AND USER MISUSE ANALYSIS (SHUMA), AND CORRECTIVE AND PREVENTATIVE ACTION (CAPA). THE DHR WAS REVIEWED AND INDICATED NO ANOMALIES WHICH COULD HAVE CONTRIBUTED THE CUSTOMER'S EXPERIENCED ISSUE. THE UNIT MET ALL SPECIFICATIONS AT THE TIME OF RELEASE. THE SERVICE HISTORY FOR THIS UNIT FOR THE PAST SIX MONTHS (DECEMBER 2012 THROUGH JUNE 2013) DID NOT IDENTIFY ANY SIGNIFICANT TREND. THE TREND FOR THE PRODUCT MALFUNCTION CODE OF ODOR/SMELLS IDENTIFIED A BREACH DUE TO THE OUS REMEDIATION EFFORT CONTAINED IN A CAPA. A REVIEW OF THE COMPLAINT HISTORY DID NOT REVEAL A SIGNIFICANT TREND FOR US DATA OVER THE PAST TWELVE MONTHS (JANUARY 2013 THROUGH DECEMBER 2013). THE FMEA REVEALED THE RISK IS AT AN ACCEPTABLE LEVEL. THE SHUMA INDICATES THE RISK IS BROADLY ACCEPTABLE FOR MODERATE LEVEL INJURY OR EXPOSURE AND AS LOW AS REASONABLY PRACTICABLE FOR MILD (LIMITED) EXPOSURE TO ODOR OR ODORANTS. CAPA INVESTIGATION AS WELL AS THE ASTM TESTING PERFORMED INDICATED OIL DEGRADATION CAN CAUSE ODOR/SMELLS FOR STERRAD® SYSTEMS. A PM2 WAS PERFORMED AND THE INVOLVED UNIT WAS LEFT IN WORKING ORDER. NO PARTS WERE AVAILABLE FOR ANALYSIS. THE ISSUE WILL BE TRACKED AND TRENDED. NO FURTHER INVESTIGATION IS NECESSARY AT THIS TIME.
AN INTERNATIONAL CUSTOMER REPORTED A "MUSHROOM/OIL SMELL" (ODOR) EMITTING FROM THE STERRAD 50 STERILIZER. THERE WAS NO REPORT OF ANY HUMAN REACTION. A FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY EVENT DATED (B)(6) 2013.
CORRECTED INFORMATION: THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY EVENT DATED (B)(6) 2013.