FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3203479 · Received July 2, 2013

Report

Report Number
2032227-2013-02714
Event Type
Malfunction
Date Received
July 2, 2013
Date of Event
June 6, 2013
Report Date
June 6, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP FAILED THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR DURING THE REWIND DUE TO AN OUT OF PHASE MOTOR ENCODER SIGNAL.

Description of Event or Problem · 1

THE CUSTOMER CALLED TO REPORT A MOTOR ERROR ALARM DURING NORMAL USE. THE CUSTOMER STATED THAT THE INSULIN PUMP WAS EXPOSED TO AN MRI SCAN ON (B)(6) 2013. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP WAS NOT DAMAGED. THE INSULIN PUMP FAILED THE DISPLACEMENT TEST, AND ALARMED MOTOR ERROR AGAIN. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
302311 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722RNAS

Patients

Seq Age Sex Outcome Treatment
1 65 YR