FDA Adverse Event
Malfunction
Summary report: N
MEGA NEEDLE DRIVER INSTRUMENT
MDR report key: 3203474
·
Received July 2, 2013
Report
- Report Number
- 2955842-2013-02395
- Event Type
- Malfunction
- Date Received
- July 2, 2013
- Date of Event
- May 30, 2013
- Report Date
- June 6, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE GRIP CLOSE CABLE WAS FOUND TO BE BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY WAS ABLE TO SPIN FREELY AND DID NOT EXHIBIT ANY DAMAGE. THE CABLE SEGMENT STICKS OUT AT THE WRIST. OTHER CABLES AT WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE BROKEN CABLE IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE CUSTOMER NOTED A BROKEN CABLE ON THE MEGA NEEDLE DRIVER INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 302753 | MEGA NEEDLE DRIVER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420194-08 | M10120210 989 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES |