FDA Adverse Event Malfunction Summary report: N

COULTER® LH 750 HEMATOLOGY ANALYZER

MDR report key: 3203267 · Received July 2, 2013

Report

Report Number
1061932-2013-01365
Event Type
Malfunction
Date Received
July 2, 2013
Date of Event
June 5, 2013
Report Date
June 5, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061574
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT. THE FSE FOUND THAT THE GREEN STRIPE TUBING HAD POPPED OFF VC12. THE FSE REPLACED THE GREEN STRIPE TUBING FROM VL46A TO VC12 WHICH RESOLVED THE LEAK. THE FSE ALSO REPLACED THE REGULATOR TUBING THAT CORRECTED THE 30 PSI LOW ERRORS. FAILURE MODE: GREEN STRIPE TUBING POPPED OFF VC12. UNIT WAS DISPLAYING LOW PSI. PER LABELING, BECKMAN COULTER URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IS NOT LIMITED TO PROTECTIVE EYEWEAR, GLOVES AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BECKMAN COULTER (BEC) STATING THAT RIGHT AT THE START OF THE OPERATOR'S SHIFT ABOUT TO RUN THE FIRST SET OF SAMPLES SHE NOTICED LEAKING FLUID IN THE COULTER LH 750 HEMATOLOGY ANALYZER. THE ANALYZER WAS ALSO GENERATING A 30 PSI TO LOW ERROR. THE VOLUME OF LEAK WAS ABOUT ONE CUP AND WAS NOT CONTAINED WITHIN THE UNIT. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES AND A LABORATORY COAT AT THE TIME OF THE EVENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS REVIEWED. THE LABORATORY'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION TO THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
302385 COULTER® LH 750 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER LH 750 NA

Patients

Seq Age Sex Outcome Treatment
1